Tel Number
180-3717-8440
Model:KJ-A1700-8L-DZ
Description:
This customized high-temperature atmosphere furnace is mainly used in college laboratories, industrial and mining enterprise laboratories.
Product number |
KJ-A1700-8L-DZ |
Furnace structure |
Double-layer shell structure, equipped with a cold air system to automatically cool the furnace shell |
Furnace material |
High-quality and high-purity alumina polycrystalline fiber curing furnace, with good insulation performance, high reflectivity and balanced temperature field |
Maximum temperature |
1700 ℃ |
Operating temperature |
≤ 1600 ℃ |
Heating rate |
Recommended ≤ 10 ℃ /min |
Heating zone size |
200*200*200mm (width * height * depth) |
Temperature Control System |
With PID adjustment and self-tuning functions, and can program 30 sections of temperature rise and fall programs; |
Heating element |
High temperature silicon molybdenum rod |
Temperature measuring element |
Type B thermocouple |
interface |
The furnace is equipped with inlet, outlet and vacuum ports, and there are float flow meters and mechanical vacuum pressure gauges on the gas path |
Sealing system |
The furnace door and roof are sealed with high temperature silicon rubber. The furnace has air inlets and outlets for gas and vacuum |
Gas line |
One nitrogen float flow meter , one hydrogen float flow meter |
Power supply |
AC220V , 50/60Hz |
rated power |
8KW |
Door opening method |
Side door |
Vacuum pump |
VRD-8 rotary vane pump pumping speed 2.2L/S |
Ultimate vacuum |
-0.1M Pa |
Exhaust gas treatment device |
The hydrogen combustion device can automatically ignite the eliminated hydrogen |
The standard configuration |
Crucible tongs a handle, a high temperature gloves 1 pair, instructions 1 parts certificate warranty card 1 parts |
Remarks |
1. Please pay special attention to: H2 is a particularly flammable and explosive gas. Its combustion temperature is 400 ℃. It will explode when the mixing ratio with air reaches 4.0%-74.2% , so safety measures must be taken when using hydrogen. 2. The maximum pressure that can withstand in the furnace is 0.03MPa , so a pressure reducing valve must be installed on the gas cylinder when gas is introduced into the furnace. 3. Before entering hydrogen, inert gas must be used to clean the furnace. The pressure inside the furnace should always be kept between 0.01-0.02 Mpa during use to prevent air from flowing back into the furnace and causing an explosion. 4. The room where the equipment is placed must be well ventilated.
5. The discharged hydrogen must be combusted indoors (the top of the furnace body is equipped with an automatic ignition device) In order to keep the pressure in the furnace chamber relatively stable during the heating process, we recommend that the equipment increase temperature rate is less than 10 ℃ /min. |
Tel Number
180-3717-8440
web@kejiafurnace.com